3000.doc Iss. 2/00
11
The channels available to add to the scanning list are designated A0 and B0 (A0…A4 , B0…B4 if the
extended scanner option is fitted). The Flashing channel may be added with Add key or deleted with
the Del key moving between the channels with the < > keys. Once a channel is added to the scan list
the channel is highlighted, pressing the OK key stores that scanning sequence.
To set the number of scan cycles and the delay between cycles select the Timer and the above menus
will be displayed. The changeable parameters will flash. To change the flashing parameter press Edit
and enter the appropriate number of Cycles 0 to 9999 or continuous. Pressing the Clear key returns to
the default setting. The following may be set Cyc - This is the number of scanning cycles any number
between 1 and 9999, continuous can also be set. NOTE if the data logger is turned on then the number
scan cycles is limit to 4000 readings. Delay - This sets the delay before the scanner cycle starts and
may be set between 00:00:00 and 59:59:59. Rate - This sets the time between each channel selection
and may be set between 00:00:00 and 59:59:59.
Example: Set Cyc to 10
Set Delay to 00:10:00
Set Rate to 00:00:40
The channels will be scanned 10 times with a start delay of 10 minutes and 40 seconds delay between
each channel selection.
The measurement display will start in the measurement Hold position Press TRIG to start the scan
cycle.
A0 Pt100 EN60741 R0 = 100.000 a = 3851 4wre
+ 100.000
°
C
-------------------------------------------------------------------------------
Scanning list edit: SL1
A0 B0
< > Add Del Quit OK
A0 Pt100 EN60741 R0 = 100.000 a = 3851 4wre
+ 100.000
°
C
-------------------------------------------------------------------------------
Cyc = Cont Delay = 00:00:00 Rate =
00:00:00
< >` Clear Edit Quit OK
A0 Pt100 EN60741 R0 = 100.000 a = 3851 4wre
+ 100.000
°
C
Input =138.500 Ohm +I Busy
Hold SL1 0/10