MicrohmmeterType DO5000 SeriesOperating InstructionsBracken Hill, South West Industrial Estate,Peterlee, Co. Durham SR8 2SW. England.Tel: +44 (0)191-5
Operating InstructionsMicrohmmeter Type DO500051.5 Specification1.5.1 Measurement Accuracy5000/50015002The accuracy is quoted at full rated current wi
Operating InstructionsMicrohmmeter Type DO500061.5.4 Temperature compensationThe accuracy of the temperature measurement is ±0.1% and uses a standard
Operating InstructionsMicrohmmeter Type DO500072 Setting up the DO50002.1 Safety informationPlease read and follow these important safety instructions
Operating InstructionsMicrohmmeter Type DO500082.2 Unpacking the instrumentWhen you unpack the DO5000, check that the following items are present befo
Operating InstructionsMicrohmmeter Type DO500093 About the DO5000This section introduces you to the features and functions of the DO5000 Microhmmeter.
Operating InstructionsMicrohmmeter Type DO5000103.4 Table Summary of basic functionsKey Symbol Description FunctionDirect function or menuSelection
Operating InstructionsMicrohmmeter Type DO500011Setting up measurement optionsCURR Current selection Selects the current to be used for measuremen
Operating InstructionsMicrohmmeter Type DO5000123.5 About the DO5000 Display screenThe liquid crystal graphic display clearly indicates the measured r
Operating InstructionsMicrohmmeter Type DO5000133.6 About the Function keysThe top row of function keys are only used to select menu options. The lowe
Operating InstructionsMicrohmmeter Type DO5000144 Setting up Measurement OptionsNote: It is possible to reset the instrument to the original default s
Operating InstructionsMicrohmmeter Type DO5000Limited Warranty & Limitation of LiabilityCROPICO guarantees this product for a period of 1 year. Th
Operating InstructionsMicrohmmeter Type DO500015DisplayThe main value display can be set to read Ω, Ω/\km or Km. From the measurement [MODE]screen se
Operating InstructionsMicrohmmeter Type DO500016For manual operation the Slow measurement mode should be selected. This will read at 2measurements per
Operating InstructionsMicrohmmeter Type DO500017Figure 4.4.1 Measurement screen with temperature compensation4.5 LimitsPress [Lim] To select the Limit
Operating InstructionsMicrohmmeter Type DO5000184.6 Data Logging FunctionPress [LOG] to select the data logging menu Select [New] to initiate a new lo
[ ] Operating InstructionsMicrohmmeter Type DO500019The following parameters will be stored when saved to memory:-PARAMETER DEFAULT VALUERange 30kOhm
Operating InstructionsMicrohmmeter Type DO5000204.9 Measure/Hold functionPress [Mes] to hold the measured value on the display. NOTE this does not in
Operating InstructionsMicrohmmeter Type DO5000215 Measuring with the DO50005.1 Connecting to the DO5000The DO5000 uses a four terminal method of measu
Operating InstructionsMicrohmmeter Type DO500022Figure 5.3.1 Diagram of different connection configurationsC1 P1 P2 C2Connection to Cable JointP1 P
Operating InstructionsMicrohmmeter Type DO5000235.4 Resistance measurementThe DO5000 should be in the measurement [STOP] mode. In this mode the measur
Operating InstructionsMicrohmmeter Type DO5000245.6 Measurement with LimitsThe measurement limits function permits the fast easy sorting of components
Operating InstructionsMicrohmmeter Type DO5000Disposal of Old ProductThis product has been designed and manufactured with high quality materials andco
Operating InstructionsMicrohmmeter Type DO5000255.7 Measurement of Inductive DevicesGENERALWhen the Settling algorithm is selected, and the ‘MEAS’ but
Operating InstructionsMicrohmmeter Type DO500026OTHER ISSUESDue to inductive effects, ‘Overange Error’, ‘Polarity Error’, ‘Excess Resistance in Curren
Operating InstructionsMicrohmmeter Type DO5000276 Accessories6.1 LeadsThe DO5000 may be used with a variety of lead sets. The following are the availa
Operating InstructionsMicrohmmeter Type DO5000286.5 Temperature ProbeFor the sensing of the temperature of the sample under test.Ordering Code Descri
Operating InstructionsMicrohmmeter Type DO500029For Technical Support Contact:Tel: +44 (0) 191 587 8718For Service and Calibration Contact:Service Dep
Operating InstructionsMicrohmmeter Type DO500030Additional Products From CropicoDIGITAL OHMMETERSRESISTANCE BRIDGESRESISTANCE DECADE BOXESRESISTANCE S
610018 Rev 11
Operating InstructionsMicrohmmeter Type DO5000English Important Safety Information 10/08/99GeneralThis instrument has been designed and tested to co
Operating InstructionsMicrohmmeter Type DO5000Table of Contents1. Introduction 11.1 Overview 11.2 Definitions and Terminology 21.3 Standard Features o
Operating InstructionsMicrohmmeter Type DO500011. Introduction1.1 OverviewThe DO5000 Microhmmeter is a high accuracy instrument designed for industria
Operating InstructionsMicrohmmeter Type DO500021.2 Definitions and TerminologyµΩmicrohmm 0.000001 OhmmΩmilliohm 0.001 OhmkΩkilohm 1000 OhmPt100 Platin
Operating InstructionsMicrohmmeter Type DO500031.3 Standard features of the DO5000 Microhmmeter include:• Direct display of resistance• Illuminated di
Operating InstructionsMicrohmmeter Type DO500041.4 Principles of measurementBLANK PAGE